6.8.3 pause monitoring session (optional)
back to index | back to monitoring the patient
Pausing a monitoring session
End users have the option to pause a monitoring session to allow for intended or non intended interruptions. This removes unnecessary or undesired monitoring time.
Step1 - Pause session
In a monitoring session main screen click “pause monitoring” .
Step2 - Resume session
A double line pause symbol will show up in the middle of the screen. To resume, the user need to click on the “resume monitoring” button.
Screen item Nr | Name | Description |
---|---|---|
1 | Pause symbol | Increased pause symbol to make clear that monitoring session is paused |
2 | Session time | The monitoring session time counter will stop and blink in grey scale |
3 | Resume button | The “Pause monitoring” button will be replace by “Resume /monitoring session”. |
Viewing a pause in a final report (original report)
A monitoring session pause will be processed and showed in the final report in the following manner.
Pause in the Estimated P2/P1 ratio variation over time
In the main graph of minute by minute estimated P2P1 ratios, the paused section of the monitoring session will be removed and a blue line with a pause symbol will be places between the last processed minute before the pause and the first processed minute after the pause.
Pause in the Monitoring results section
In the minute-by-minute section of the average processed surrogate ICP waveform and derived parameters, the paused section will be indicated by a solid blue rectangle with a pause symbol.
Minutes fully within the pause period will not display waveform shape information or derived parameters.
Minutes partially overlapping with the pause will show the start and end times of the pause. If there is sufficient information in those minutes for parameter derivation, the average pulse waveform and derived parameters will be displayed.
Pause in the Additional Information section
In the Additional Information report section, all minutes fully contained within the pause will be removed, except for the B4C signal over time graph. A blue line with a pause symbol will be placed between the last processed minute before the pause and the first processed minute after the pause.