6.8 monitoring the patient
back to index | back to instructions for use
Additional help for monitoring the patient
Step 1 - position patient
Refer to 6.5 positioning patient
Step 2 - start monitoring session
Refer to 6.6 start monitoring session
Step 3 - position sensor on patient
Refer to 6.7 positioning sensor on patient
Step 4 - monitor patient
Once monitoring has started, a screen is shown in the mobile app, as shown in the figure below, where some information is presented to aid monitoring.
ATTENTION - For optimal monitoring session quality, it is recommended that the patient remains at rest and as still as possible.
PRECAUTION - The patient position or movement and the BcSs-PICNIW-1000 and BcSs-PICNIW-2000 sensors is known to affect the signal being recorded. As a result, if multiple recording sessions are obtained on the same patient, consistent positioning of the patient and device will lead to more comparable recordings.
Monitoring screen features
Function | Description | |
---|---|---|
1 | Edit patient info | Access the patient information screen to view and edit details. |
2 | morphology tab | live waveform screen displays the last 5 seconds of the monitoring session. |
3 | trend tab | displays the full or partial waveform stream of the monitoring session. |
4 | P2P1 tab | A tab that displays the estimated P2P1 ratio, either minute by minute or in 20-second intervals. |
5 | TTP tab | A tab that displays the normalized Time-to-Peak, either minute by minute or in 20-second intervals. |
6 | Observations | Opens the notes panel to add clinical notes to the monitoring session |
7 | Partial Report | A partial report access button, where the two digits before the “m” indicate the number of minutes from the monitoring session that will be included in the report. |
8 | external | External battery charge level in % (only relevant to BcSs-PICNIW-1000 sensor model) |
9 | Bluetooth signal strength | Bluetooth signal strength status. Green (good), yellow (regular) and red (bad) |
10 | internal | Internal battery charge level in % |
11 | UDI | Sensor’s Unique ID number |
12 | Monitoring time | Time elapsed since start of monitoring session |
13 | UMI | Unique Monitoring Session ID |
14 | End monitoring | End monitoring session button |
15 | Pause monitoring | Pause monitoring session button (Please refer to section 6.8.3 pause monitoring session (optional) ) |
Step 5 - interpret surrogate ICP waveform
Step 6 - finalize monitoring session
To end the monitoring session, click "End Monitoring." A popup window will appear, asking you to confirm. Click "Save" to finalize.
After the monitoring session ends, you will be directed to a confirmation screen. Here, you can fill in or edit patient data before generating the final report. To access a partial report, click the expand/zoom button on the right side.
Once everything is confirmed, click "Confirm" to submit the data to the B4C System and view the report in the mobile app. Alternatively, you can click "Postpone" to save the data and submit it later.
Step 7 - final monitoring session report
If you choose to confirm the patient data, the next screen will display the monitoring report.
The intermediate report screen contains interactive elements described bellow:
Report status:
PARTIAL = parts of the submitted data still needs to be processed.
FINAL = all monitored data has been processed.
Monitoring status:
IN PROGRESS = monitoring in progress, the report being viewed is an intermediate report.
ENDED = monitoring has ended.
refresh button = is active when pending data has been processed and made available.
repeat button = you can choose to repeat the monitoring session reusing patient information.
Option to export and open PDF
Export PDF - option to download the report in PDF format
Open PDF - option to open a downloaded PDF report
If you choose to open PDF, a file of the final report is saved on your local mobile device. When you click “Open”, the mobile device will prompt you to use an installed application to read the PDF file or ask for one if none have been installed. Select the application you normally use to view the report.
ATTENTION - Avoid downloading reports in PDF format on your mobile device. Please read section 6.9.1 b4c report (original) for a detailed view on the B4C System’s report.
back to top | go to b4c report
Copyright © 2020 - 2024 Braincare Desenvolvimento e Inovação Tecnológica S.A. All rights reserved.