6.8.1 interpret surrogate ICP waveform

back to index | back to monitoring the patient


Typical ICP waveform shapes

A satisfactory waveform reading may present any of the typical ICP shape characteristics below according to the patient’s clinical condition (from normal to pathological)

image-20240825-022931.png
ICP  waveforms. Source: Nucci CG, et al. Acta Neurochir (Wien). 2016. Intracranial pressure wave morphological classification: automated analysis and clinical validation.

back to top


Suggestive reference for ICP waveform shapes

As a suggestive reference, the waveforms below are non invasive ICP waveforms that can be shown with the noninvasive ICP sensor.

WARNING - The appearance of a normal waveform morphology does not eliminate the possibility of elevated intracranial pressure or irregularity in brain compliance.

image-20240825-023027.png

ATTENTION - The waveforms above are examples of typical noninvasive ICP waveforms in different clinical settings. You may see variations in the ICP waveform, but they should always follow a similar pattern.

back to top


Typical session duration

The monitoring session should be as long as required to assess the ICP waveform. Typical sessions are of 15+ minutes but the final duration is ultimately defined by a trained professional. The session should be finalized upon patient request due to discomfort or decision by the primary caregiver.

back to top


ICP pulse waveform

The ICP pulse waveform is an alternative to assess certain ICP characteristics, irrespective of the absolute values that are typically captured by an invasive method .

The ICP waveform contains characteristics that can be mathematically analyzed such as:

  • Wave peaks: P1 (percussion wave, due to arterial pressure being transmitted from the choroid plexus to the brain ventricles), P2 (tidal wave, related to brain compliance) and P3 (dicrotic wave);

  • Absolute amplitude of wave peaks: dP1 (relative amplitude of percussion wave), dP2 (relative amplitude of tidal wave), dP3 (relative amplitude of dicrotic wave);

  • Absolute lags in milliseconds of wave peaks: LT (considered the initial timestamp of a pulse), L1 (lag for percussion wave), L2 (lag for tidal wave), L3 (lag for dicrotic wave), Lx (variable lag);

  • Absolute curvature of wave peaks: Curv1 (percussion wave curvature), Curv2 (tidal wave curvature), Curv3 (dicrotic wave curvature).

Under normal ICP conditions, the relative amplitudes of the three peaks are related as follows: P1 > P2 > P3 (Box A in Figure above). In conditions with decreased brain compliance and rising ICP, the pulse waveform morphology gradually changes and certain indicators, such as the estimated P2/P1 ratio eventually increases.

back to top


Interpretation warnings

WARNING - The BcSs-PICNIW-1000/BcSs-PICNIW-2000 Sensor does not replace a comprehensive clinical evaluation. The waveform output should always be evaluated by the clinician in conjunction with other clinical parameters or analyses.

back to top