ATTENTION - make sure the steps below have been completed:

Table of Contents
minLevel1
maxLevel7

...

...

Step 1 - start session

From the main screen click on new session (1) button to start monitoring session.

Tip

TIP - For login instructions and pairing sensor please read section 6.3 loging log into app and pairing sensor.

...

Step 2 - fill in patient data (optional)

At the beginning of the monitoring session, an optional step is to fill patient data. This information will be used and recorded in the final report. Fill in all data correctly and click on “start monitoringpositioning(1) as shown in Figure 6.6-2below.

Tip

TIP - For faster monitoring, you can choose to skip this step, but for the report to be generated, you will have to fill in the patient information at the end of the monitoring session.

...

Categorize session with tags (optional)
Info

ATTENTION - report classification via tags must be activated by your organization before you can use it.

If activated, you can classify your monitoring session with predefined tags to facilitate report filtering in the report listing screen.

Click on the report category (plus) button.

...

A popup window will appear with the available classification tags your organization made available in advance. Choose one or more tags to classify the monitoring session.

...

Info

ATTENTION - Classification tags are defined centrally by your organization such as available tags as well as minimum to maximum amount of required and allowed tags respectively.

Tip

TIP - You can add classification tags in all patient information forms that are available at the start, middle and end of a monitoring session.

...

Tip

TIP - if activated for your corporate account you can tag your monitoring session reports. Please refer to 6.6.2 tag monitoring sessions (optional)

...

Step 3 - verify positioning (optional)

Once satisfied with the pulse morphology displayed on the app’s sensor positioning screen Figure 6.6-5, click the start monitoring button (1).

Info

ATTENTION - For instructions on how to position the sensor and verify the waveform technical quality please read 6.5 positioning sensor and verifying signal please refer to 6.6 position sensor section of this user guide.

...

Step 4 - monitor patient

Once monitoring has started, a screen is shown in the mobile app, as shown in Figure 6.6-6the figure below, where some information is presented to aid monitoring. Table 6.6-1 describes each screen field.

Info

ATTENTION - It is recommended for good monitoring session quality that the patient be kept at rest and as still as possible. Please refer to Section 6.4 positioning the 5 position patient

Note

PRECAUTION - The patient position or movement and the BcSs-PICNIW-1000 | BcSs-PICNIW-2000 sensor is known to affect the signal being recorded. As a result, if multiple recording sessions are obtained on the same patient, consistent positioning of the patient and device will lead to more comparable recordings.

...

Monitoring screen features

No.

Function 

Description

1

Back

End monitoringEdit patient info

Access to patient information screen for visualization and edit

2

See morphology tab

Directs to signal live waveform screen showing last 5 seconds of monitoring session

3

Edit patient data

Opens a form for filling in and editing patient data, as well as including observations related to the monitoring session 

4

See trend tab

Directs to trend screen tab that shows the full or partial monitoring session waveform stream

4

P2P1 tab

tab that shows minute by minute derived parameter of estimated P2P1 ratio (please refer to

5

UDI

Paired sensor identifierTTP tab

tab that shows minute by minute derived parameter of normalized time-to-peak (please refer to

6

UMI

Monitoring session identifierObservations

Opens the notes panel to add clinical notes to the monitoring session

7

Battery charge

Percentage level of internal battery (right) and external battery (left)

8

Partial Report

Partial report access button where the two digits before the “m” informs how many minutes of the monitoring session will be included in the report

8

external

External battery charge level in % (only relevant to BcSs-PICNIW-1000 sensor model)

9

Bluetooth signal strength

Bluetooth signal strength status. Green (good), yellow (regular) and red (bad)

9

10

internal

Internal battery charge level in %

11

UDI

Sensor’s Unique ID number

12

Monitoring time

Time elapsed since start of monitoring session

10

13

UMI

Unique Monitoring Session ID

14

End monitoring

End monitoring session button

11

01m Report

Partial report access button where the two digits before the “m” informs how many minutes of the monitoring session will be included in the report

12

Add Notes

Opens the observation add panel

...

15

Pause monitoring

Pause monitoring session button (Please refer to section 6.6.3 pause monitoring session (optional) )

Step 5 - interpret

...

surrogate ICP waveform

A satisfactory waveform reading may present any of the typical ICP shape characteristics below (Figure 6.6-3) according to the patient’s clinical condition (from normal to pathological)

...

Include Page
6.6

...

Warning

WARNING - The appearance of a normal waveform morphology does not eliminate the possibility of elevated intracranial pressure or irregularity in brain compliance.

 

...

Info

ATTENTION - The waveforms above are examples of typical noninvasive ICP waveforms in different clinical settings. You may see variations in the ICP waveform, but they should always follow a similar pattern.

The monitoring session should be as long as required to assess the ICP waveform. Typical sessions are of 15+ minutes but the final duration is ultimately defined by a trained professional. The session should be finalized upon patient request due to discomfort or decision by the primary caregiver.

The ICP pulse waveform (Figure 6.6-9) is an alternative to assess certain ICP characteristics, irrespective of the absolute values that are typically captured by an invasive method .

...

The ICP waveform contains characteristics that can be mathematically analyzed such as:

  • Wave peaks: P1 (percussion wave, due to arterial pressure being transmitted from the choroid plexus to the brain ventricles), P2 (tidal wave, related to brain compliance) and P3 (dicrotic wave);

  • Absolute amplitude of wave peaks: dP1 (relative amplitude of percussion wave), dP2 (relative amplitude of tidal wave), dP3 (relative amplitude of dicrotic wave);

  • Absolute lags in milliseconds of wave peaks: LT (considered the initial timestamp of a pulse), L1 (lag for percussion wave), L2 (lag for tidal wave), L3 (lag for dicrotic wave), Lx (variable lag);

  • Absolute curvature of wave peaks: Curv1 (percussion wave curvature), Curv2 (tidal wave curvature), Curv3 (dicrotic wave curvature).

Under normal ICP conditions, the relative amplitudes of the three peaks are related as follows: P1 > P2 > P3 (Box A in Figure 6.6-4 above). In conditions with decreased brain compliance and rising ICP, the pulse waveform morphology gradually changes and certain indicators, such as the estimated P2/P1 ratio eventually increases.

Warning

WARNING - The BcSs-PICNIW-1000 Sensor does not replace a comprehensive clinical evaluation. The waveform output should always be evaluated by the clinician in conjunction with other clinical parameters or analyses.

Warning

WARNING - The B4C System has not been validated to be a substitute for invasive ICP monitoring methods in any cases.

Warning

WARNING - The medical professional is responsible for determining the additional clinical parameters or analyses required in addition to the BcSs-PICNIW-1000 Sensor waveform to make a clinically informed decision.

Warning

WARNING - The appearance of a normal waveform morphology does the eliminate the possibility of elevated intracranial pressure or irregularity in brain compliance.

Warning

WARNING - The clinician is reminded to assess the waveform morphology in real-time during a monitoring session. The B4C report is available for the clinician’s convenience to review associated waveform parameter estimates and other clinical observations entered in a past monitoring session

...

.1 interpret surrogate ICP waveform
6.6.1 interpret surrogate ICP waveform

Step 6 - finalize monitoring session

To end the monitoring session click on end monitoring. A popup window will appear asking you to confirm the end of the monitoring session. Click on save.

Once the monitoring session is complete you will be directed to a confirmation screen. Here you can fill in or edit patient data before the final report is generated. You can also access a partial report by clicking on the expand/zoom button on the right side.

Once confirmed, you can click “Confirm”(1) which will submit the data to the B4C System and allow you to view the report in the mobile app, or you can click “Postpone”(2) and then submit the data in the future.

...

Step 7 - final monitoring session report

If you choose to confirm the patient data, the next screen will be the monitoring report view. The screen elements are described bellow:

...

If you choose to open PDF, a file of the final report is saved on your local mobile device. When you click “Open”, the mobile device will prompt you to use an installed application to read the PDF file or ask for one if none have been installed. Select the application you normally use to view the report.

Info

ATTENTION - Avoid downloading reports in PDF format on your mobile device. Please read section 6.7 b4c report for a detailed view on the B4C System’s report.

...