Unable to render embedded object: File (logo_brain4care.png) not found.

Skip to end of metadata
Go to start of metadata

You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 2 Next »

back to index | back to multiparameter monitor

In this pairing method, the B4C System mobile app will make the connection between the dongle and the sensor. A major benefit of this method is the ability to view the sensor waveform both on the mobile app and patient monitor and also the ability to generate a monitoring session report (please view section 6.9.1 b4c report (original) for details on the generated report).


Step 1 - Connect receiver (dongle)

Connect the receiver/dongle (nr.1) to the invasive pressure input on the patient monitor (nr.2) and make sure.

ATTENTION - The dongle requires a minimum power supply of 5V. Typically this is supplied through the ABP plug. When this is not available a charging cable connected to the dongle (nr.3) is required. The required cable is a Micro-Usb on the dongle’s side and normal Type A USB on the monitor side.


Step 2 - Log into app and pair sensor

Log into app and pair sensor as explained in section 6.3 logging into app

Step 3 - Initiate dongle pairing in mobile device

After pairing the sensor, in the application, click “Pair Receiver.” A list of available receivers will appear.

Check the receiver’s UDI on the label opposite to the button… (see item 1 in figure 6.8-4)

…and then select it from the list (item 1 in figure 6.8-5). Some receivers may not be available to be clicked, represented in gray, as they are not active, locked, do not belong to the organization (corporate account), or are awaiting activation (if your sensor is grayed out, check with your clinical engineering team).

Step 5 - Verify matching devices in app

After selecting the dongle, the pairing intelligence mode between sensor and receiver will start, with a visual LED confirmation on both devices. Visually confirm the pairing with the physical receiver (Figure 6.8-6), with the receiver image in the mobile app. Both should be flashing matching color and blinking patter. Click “Confirm” to continue pairing (item 1 in figure 6.8-6). If the pattern is not identified or you want to change the receiver, simply click on “Reselect”.

Once the connection with the dongle is confirmed, you will be directed to the screen on Figure 6.8-7, where the sensor, receiver and application are paired. The same pairing intelligence method is done with the sensor, receiver, and mobile application screen by flashing in the same randomly generated color and frequency pattern.

Step 6 - Confirm dongle pairing

Verify that all three devices have the same flashing pattern and click “Confirm” (item 1 Figure 6.8-7). If the pattern is not identified or you want to change any of the devices, just click “Reselect”. (Figure 6.8-7)

The sensor and receiver status LEDs will flash green three times as soon as pairing is confirmed, and a success window will appear in the application.

Step 7 - Verify signal on patient monitor

At this point, the multiparameter monitor should start receiving the signal from the waveform. For best viewing, set the patient monitor to automatic scaling. If the signal is not found on the multiparameter monitor, turn off the sensor, unplug the receiver, and return to step 1.

Remark: If in addition to viewing on the patient monitor you also want to have a generate the report in the mobile application, click on “Start monitoring” and proceed to section 6.8 monitoring the patient . If you just want to monitor on the multiparameter monitor, click “Cancel”.

WARNING - Do NOT attempt to calibrate BcSs-PICNIW-1000 Sensor with your patient monitor. This may result in display of a misleading ICP value shown in the patient monitor that can be misinterpreted when making critical clinical decisions.

WARNING - Make sure NO ICP value is shown in your patient monitor when the BcSs-PICNIW-1000 Sensor is connected. Make sure that the default question marks (? /?) or dashes (- / -) are shown instead.

ATTENTION - Only use with a compatible monitors and DO NOT calibrate the ICP value (mmHg)

back to top

  • No labels