To disinfect the sensor generously apply 70% ethanol alcohol to a sterilized gauze and wipe each component according to the following instructions:
ATTENTION - Do NOT sterilize in an autoclave. This will damage the equipment.
ATTENTION - Do NOT press the pin on the green base of the sensor during sterilization to avoid damaging the sensor.
Step 1 - turn off sensor
Before you begin, make sure the sensor is turned off verifying that the power switch set to “O”.
Step 2 - disassemble sensor if needed
Disconnect each component from the sensor set and clean them separately.
Step 3 - apply alcohol
Apply ethyl or isopropyl alcohol (70-90%) to sterile gauze towels and gently wipe each component (sensor, battery, headband, and turnbuckle).
Step 4 - let it dry
Wait for the components to dry for a few minutes and then store them in the storage case.
Disinfection must be done before and after each monitoring session to prevent contamination between monitoring sessions and storage. Components that have not been properly disinfected carry the risk of contamination.
ATTENTION - Disinfect the sensor and headband with 70% ethanol alcohol, or other agents according to your healthcare facility’s disinfection protocols. This must be done before and after any use on patients.