4.1

...

sensor

BcSs-PICNIW-2000 sensor

Sensor Technical Specifications

Sensor

Value

Supply Voltage

3.7 Vdc

Power 1

MédiaAverage: 0.120W | MáximaMaximum: 0.6 W

Dimensions (H x W x D)

50 x 35 x 30 mm

Weight

64g

...

Info

ATTENTION - If the BcSs-PICW-2000 sensor model is stored in locations of up to 40 ºC, it is recommended that the sensor be used in a location with a temperature within the temperature operating range and wait until the sensor cools down. If the operating environment is at 20 ºC, the sensor takes approximately 6 minutes to cool down and reach the operating temperature (35 ºC).

...

Info

ATTENTION - Only use Brain4care authorized accessories and parts

back to top

...

4.5

...

chargers

Charger of BcSs-PICNIW-2000 sensor

...

External battery charger settings

Parameter

Specification (typical)

Battery voltage regulation

4.2 V

Current during fast charge

0.979 A

Pre-charge to fast-charge transition threshold

3 V

Termination current

96.8 mA

Timers

Pre-charge: 1800s

Safety timer: 18000s

Battery temperature regulation

Charging suspension: 45°C

Charging resumption: 40°C

back to top

...

4.6 mobile device

The brain4care app is compatible with mobile devices that meet the following requirements:

  • Android: Version 7.0 or higher, with a minimum of 3 GB of RAM and Bluetooth® 4.2 or 5.0.

  • iOS: Version 15 or higher, with a minimum A12 processor and Bluetooth® 4.2 or 5.0.

The app connects to the supplied sensors (BcSs-PICNIW-1000 or BcSs-PICNIW-2000) and optionally to the BcAc-DG-1000 receiver. It enables real-time visualization of the sensor waveform and cloud-processed waveform information, including derived parameters.