Step 1 - open app

Open the brain4care app on you mobile device. You will be directed to the APP login screen for Standard Login or the active user selection screen for Trusted Device Login

...

Info

ATTENTION- Make sure the app is already installed. If not please read section https://brain4care.atlassian.net/wiki/spaces/B4CSENSORGUIDE29/pages/2853210148/5+equipment+installation#5.0-mobile-app-installation

Standard login

...

Step

...

On the home screen, log in using the ID option (1), followed by the username or ID + password(2), and press login (4).

...

Step 2 - select a corporate account (optional)

...

2

...

- log into

...

Enter the corporate password(1) and click LOGIN. If the account you want does not appear, contact your clinical engineering team to request access to that account.

...

Trusted device login

Info

ATTENTION - this login option is available for registered TRUSTED devices. Contact your clinical engineering team to enable this on the device you are using. (Refer them to <link> of this userguide).

Step 1 - select user profile

Choose your user profile from the list of active users in your corporate account.

...

Step 2 - enter PIN (optional)

Enter your individual PIN. This is only required of your clinical engineering team has defined that individual user PINs are required for the trusted devices.

...

app

Child pages (Children Display)
allChildrentrue

...

Step 3 - first screen after login

After login a main screen will be shown with three top sections:

...

Icon

Status

Description

Report available in the cloud

Indicates that the report is in the cloud. In this case, the mobile device needs to be ONLINE to access the report.

Report locally available

Indicates that the report was downloaded on the mobile device, i.e., it is available offline.

6.3.2 - pairing the sensor

Step 1 - turn on the sensor

To turn on the sensor to initiate pairing procedure.

Turn on sensor using power switch, setting it to “I” position.

Image Removed

Turning ON sensor BcSs-PICNIW-1000

Make sure you are not trying the press the MODE button. This is used to pair the sensor with a dongle (please see section https://brain4care.atlassian.net/wiki/spaces/B4CSENSORGUIDE29/pages/2853211499#6.8.1-manual-pairing if you are using a dongle)

Info

If the switch is not in its final position, the internal pin protector may interfere with adequate pin movement, decreasing sensor sensitivity, or even causing misreadings.

Verify if sensor is on via LED indication. If not, long press the on/off button util the LED is on.

Image Removed

Turning ON the sensor BcSs-PICNIW-1000

Info

ATTENTION - The maximum operating temperature of the BcSs-PICW-2000 sensor model is 35°C. Therefore, if it is stored in a location of up to 40°C, it is recommended to use the sensor in a place with a temperature within the operating range (0 to 35°C) and wait until the sensor cools down. If the operating environment is at 20°C, the sensor takes approximately 6 minutes to cool down and reach the operating temperature (35°C).

Step 2 - start sensor pairing

Click on the (plus) connect button (1). The app will scan for any close by sensors that are ON.

...

Step 3 - confirm the sensor

A list of sensors will appear on your sensor pairing pop-up screen (1). The closest sensor will be selected automatically. Verify if the sensor UDI (Unique Device ID) on the sensor, opposite the power buttons, is the same that has been selected from the sensor list in the mobile app.

Tip

TIP - Some sensors may not be available for selection (dimmed) because they are blocked or not active for your organization.

...

.