5.1 mobile app installation
The brain4care app can be downloaded and installed from the Google Play Store or Apple App Store by following these steps:
Android (Google) | iOS (Apple) | ||||
Step 1 - app store: On your mobile device, open App Store: Open the app store for your operating system (iOS or Android). | |||||
Step 2 - Search for the app: Tap the search bar and enter "brain4care" to find the app. | |||||
Step 3 - app selection: Select the brain4care app from the search results. | |||||
Step 4 - installation: On the next screen, click INSTALL or GETInstall app (INSTALL for Android or GET for iOS). | |||||
Step 5 - Open the app: After downloading and installing the app, click OPEN.app after successful download and install | |||||
note | |||||
CAUTION - Make sure your mobile device is configured with a strong password; do not use it in root mode, nor with unverified apps installed | |||||
Step 6 - GPS permission: To ensure proper functioning of the B4C System, you must Enable GPS for to enable location access for the device. | |||||
Step 7 - make sure Bluetooth is active: To use the brain4care system, ensure Bluetooth is turned on. If it’s not, enable it in your device settingsActivate Bluetooth to enable communication with sensors and dongle. | |||||
Step 8 - Access corporate access account : To use the brain4care app, you must have a brain4care account. to use mobile device | |||||
Your corporate account manager must create and enable your user account for the application and system. For details, consult technical support or refer to the Portal Manual.
|
...
Info |
---|
ATTENTION - It is recommended to keep the original packaging in case the equipment needs to be transported. Misuse of the product will void the warranty. |
...
Location and Setup
Operating Temperature: Ensure it is within the recommended range.
Brightness: Choose an area a location with appropriate lighting.
Power Outlets: Ensure availability if using a multiparameter monitor that requires power.
Space: Allow sufficient enough room for both professionals and patients to move comfortably.
Environmental Conditions
Climate Control: The system should be used in a climate-controlled environment
...
. Maintain temperature, humidity, and pressure within the
...
limits specified in Section 4.2 -
...
Transportation and Storage: Ensure conditions remain within the specified temperature, humidity, and pressure
...
ranges during transportation or storage, as outlined in Section 4.2 - Environmental.
Electromagnetic Compatibility
Installation Environment: Install and use the equipment in an electromagnetic environment as described in
...
RF Communication Equipment: Maintain a distance from portable and mobile RF communication equipment
...
to avoid interference, as specified in
...
Coastal Environment
Salt Spray: In coastal cities,
...
use air conditioners and dehumidifiers to
...
Ensure the equipment, particularly the sensors, is not exposed to direct sunlight.
For accurate monitoring, the patient should remain still during the session.
...
mitigate damage from salt spray.
Warning |
---|
WARNING - Do not leave the equipment near flammable chemicals such as alcohol, benzene, etc. Should they come in contact with electrical parts, this may result in fire or internal short circuit. The same is true with the use of some flammable disinfectants. |
...
5.3 power supply
...
BcSs-PICNIW-2000 sensor
...
Features a single internal battery and is charged via an inductive charger.
The sensor cannot be used while charging.The
BcSs-PICNIW-1000 sensor
...
Operates with two lithium-ion batteries:
one internal
...
one external removable battery.
The internal battery provides sufficient power to keep the sensor operational while the external battery is being charged separately. For charging instructions please go to 6.11 charging sensor battery section.
...