ATTENTION - Do NOT press the pin on the green base of the sensor during sterilization to avoid damaging the sensor.

Step 1 - turn off sensor

Before you begin, make sure the sensor is turned off verifying that the power switch set to “O”.

Figure 6.1-1a Turning off the sensor BcSs-PICNIW-1000

Before you begin, make sure the sensor is turned off by checking if the LED is turn-on. If yes, keep pressed util the LED is turn-off

Figure 6.1-1b Turning off the sensor BcSs-PICNIW-2000

Step 2 - disassemble sensor if needed

Disconnect each component from the sensor set and clean them separately.

Figure 6.1-2a Detaching components BcSs-PICNIW-1000 sensor

image-20240206-220737.png

Figure 6.1-2b Detaching components BcSs-PICNIW-2000 sensor

Step 3 - apply alcohol

Apply ethyl or isopropyl alcohol (70-90%) to sterile gauze towels and gently wipe each component (sensor, battery, headband, and turnbuckle).

Step 4 - let it dry

Wait for the components to dry for a few minutes and then store them in the storage case.

...