Info |
---|
ATTENTION - Do NOT press the pin on the green base of the sensor during sterilization to avoid damaging the sensor. |
Step 1 - turn off sensor
Before you begin, make sure the sensor is turned off verifying that the power switch set to “O”. Figure 6.1-1a Turning off the sensor BcSs-PICNIW-1000 | Before you begin, make sure the sensor is turned off by checking if the LED is turn-on. If yes, keep pressed util the LED is turn-off Figure 6.1-1b Turning off the sensor BcSs-PICNIW-2000 |
Step 2 - disassemble sensor if needed
Disconnect each component from the sensor set and clean them separately.
Figure 6.1-2a Detaching components BcSs-PICNIW-1000 sensor | Figure 6.1-2b Detaching components BcSs-PICNIW-2000 sensor |
Step 3 - apply alcohol
Apply ethyl or isopropyl alcohol (70-90%) to sterile gauze towels and gently wipe each component (sensor, battery, headband, and turnbuckle).
Step 4 - let it dry
Wait for the components to dry for a few minutes and then store them in the storage case.
...