6.7.1. sensor placement instructions

Assessing the braincare brain4care waveform

The best position for ICP waveform monitoring is a resting supine position at 30 degrees with the chin and sternum aligned (Figure 6.7-1) and the head should not be tilted. The patient must be resting and avoid any movements since that can interfere with the ICP waveform reading.

...

Info

ATTENTION - Trained Personnel are expected to recognize proper waveform quality.

...

Warning

WARNING - Do NOT attempt to calibrate BcSs-PICNIW-1000 Sensor with your patient monitor. This may result in display of a misleading ICP value shown in the patient monitor that can be misinterpreted when making critical clinical decisions.

Warning

WARNING - Make sure NO ICP value is shown in your patient monitor when the BcSs-PICNIW-1000 Sensor is connected. Make sure that the default question marks (? /?) or dashes (- / -) are shown instead.

Info

ATTENTION - Only use with a compatible monitors and DO NOT calibrate the ICP value (mmHg)