6.1 dissinfecting sensor

back to index | back to instructions for use

To disinfect the sensor, generously apply 70% ethanol alcohol to a sterilized gauze and gently wipe each component of the sensor with the gauze, ensuring even coverage. Allow the sensor to air dry completely before reuse.

ATTENTION - Do NOT sterilize in an autoclave, as this will damage the equipment.

ATTENTION - Do NOT press the pin on the green base of the BcSs-PICNIW-1000 sensor, as this could damage the sensor.


Step 1 - turn off sensor

Before starting, ensure the sensor is turned off by checking that the power switch is set to “O”.

Figure 6.1-1a Turning off the sensor BcSs-PICNIW-1000

Before you begin, ensure the sensor is turned off by checking if the LED is on. If it is, press and hold until the LED turns off.

Figure 6.1-1b Turning off the sensor BcSs-PICNIW-2000

back to top


Step 2 - disassemble sensor if needed

Disconnect each component from the sensor set and clean them individually.

Figure 6.1-2a Detaching components BcSs-PICNIW-1000 sensor

Figure 6.1-2b Detaching components BcSs-PICNIW-2000 sensor

back to top


Step 3 - apply alcohol

Apply ethyl or isopropyl alcohol (70-90%) to sterile gauze towels and gently wipe each component: sensor, battery, headband, and turnbuckle.

back to top


Step 4 - let it dry

Wait for the components to dry for a few minutes before storing them in the storage case. Disinfection should be performed before and after each monitoring session to prevent contamination. Proper disinfection is crucial to avoid the risk of contamination from inadequately cleaned components.

ATTENTION - Disinfect the sensor and headband with 70% ethanol alcohol, or other agents as per your healthcare facility’s disinfection protocols, before and after each use on patients.

back to top


Step 5 - visual inspection

If the device is not visually clean or shows unacceptable deterioration after cleaning, either repeat the cleaning steps or safely dispose of the device.

back to top | go to assembling sensor