What are the key security measures built into your system?

  • Segregated databases for each client

  • Dedicated and separate AWS infrastructure per client that handles PHI data.

  • Private public encryption key schema for PHI data where B4C cannot access it. Only clients in an active user session can do ti via dedicated AWS infrastructure.

  • Client defined auto logoff policy for user session inactivity

  • Client defined auto user suspension policy for inactivity period

  • Client defined password policy

  • Client defined data purge policy

  • Full system audit log for over 30 user operation

What are the resilience features built into your system?

-explain how the system is designed to specify the level of resilience (independent ability to function) if communication between one component and other component of the system is disrupted. Describe potential disruption that could be considered significant.

-explain how the system is designed to be resilient to possible cyber incident scenarios such as network outages, Denial of Service, excessive bandwidth usage by other products on the same network, disrupted Quality of Service, and/or excessive jitter

-explain how the system is designed to be resilient to possible noise, if applicable
How is the system designed to be resilient to potential noise interference?

Our system is meticulously engineered to ensure high resilience and reliable performance, even in the presence of noise and environmental interferences that could affect electronic components and data communication. Here is how we address noise mitigation across different parts of our system:

  1. Sensor Resilience: Sensors are critical to our system's operation and are rigorously tested to ensure their resilience to noise. They have undergone comprehensive Electromagnetic Compatibility (EMC) testing in accordance with the IEC 60601-1-2 standard, demonstrating that they are not susceptible to noise interference from other devices, nor do they generate noise that could affect other systems. These tests ensure that the sensors maintain their functionality and accuracy, even in environments with high levels of electronic noise.

  2. Communication with Mobile Devices: Communication between the sensors and mobile devices is established using Bluetooth 5.0, chosen for its advanced noise mitigation capabilities. In dense environments, where many devices operate on the 2.4 GHz frequency, both Bluetooth and Wi-Fi connections can be impacted. However, Bluetooth 5.0 and Wi-Fi incorporates robust techniques such as channel hopping and other mitigation strategies to maintain stable and reliable communication. These techniques are detailed in our wireless coexistence document, which describes how our system minimizes interference and maintains the integrity of transmitted data.

In summary, our system leverages a combination of tested and proven hardware, advanced communication technologies, and robust data transmission protocols to ensure that noise interference is minimized and that system operation remains stable and reliable, regardless of environmental conditions.








Do you encrypt data at rest?

YES. All PHI (protected health information) is encrypted at rest. This data is encrypted at source using the public key from a client specific public-private encryption key schema. Only our clients, in an active user session, can access PHI data that is served via a dedicated client AWS infrastructure that brain4care does not have access to.

Where are your servers located?

For US customers, all servers are located in the US territory.

How to you backup data?

All data is stored in our HIPAA certified cloud infra-structure with daily backups.

Do you support Single Sign-On (SSO)?

Yes. The B4C System has SSO capabilities to integrate with customers AD (Active Directories) to allow for seamless and secure system login. This functionality is included via joint brain4care and client IT team project.

Do you support EMR integration?

Yes. The B4C System has secure APIs to integrate with customers EMR (Electronic Medical Record) to allow for seamless and secure clinical information flow. This functionality is included via joint brain4care and client IT team project.

Do you have a machine readable SBOM of all software used in your system?

Yes. Such request should be sent to security@brain4.care or via requests using our help.brain4.care portal.

How do I recover device configuration?

Only brain4care authorized representatives have access to and are allowed to view, manage and change device configurations. In B4C System’s case, the above mentioned configuration is for the sensor itself.

How are forensic log files collected and managed?

-describe how forensic log files are collected including how, where, what format they are located, stored, recycled, archived, and how it can be read by automated security analysis software

Is there an anticipated cyber-security end of support and end of life?

No. brain4care is fully committed to cyber-security coverage throughout the period of any user actively using the B4C System