CONTRAINDICATION - Do NOT use this device on patients without cranial bone flap or have extensive scalp injury, bone defects (sections of skull missing), or any other conditions that deems to be a contraindication to performing monitoring using the brain4care device.
The best optimal position for ICP waveform monitoring is at with the patient in a resting supine position, angled at 30 degrees, with the chin and sternum aligned (Figure bellowsee figures below). The Ensure the head should is not be tilted . The patient must be resting and avoiding any movements since it can interfere and the patient remains still to avoid interference with the ICP waveform reading.
Note |
---|
PRECAUTION - The patient position or movement and the BcSs-PICNIW-1000 and BcSs-PICNIW-2000 sensors is known to affect the signal being recorded. As a result, if multiple recording sessions are obtained on the same patient, consistent positioning of the patient and device will lead to more comparable recordings. |