Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

back to index | back to b4c report

Excerpt
namereport simplified

The simplified b4c report sections are:

Anchor
top
top

Table of Contents
minLevel2
maxLevel3
outlinefalse
styledefault
typelist
printabletrue

The report generated by the B4C System has five sections that organize all information acquired and processed for a monitoring session.

Info

ATTENTION - Refer to 12 technical documentation for further details regarding the analytical software and derived parameters calculation.

...

image-20240815-160840.pngImage Added

back to top


SECTION 1 - Summary

The first section is a summary of the monitoring session information including patient identification, monitoring session identification, and patient clinical context.

Header (patient Identification and context):

This section repeats itself as a header of every report page

...

image-20240815-161310.pngImage Added

this sub-section contains all information to uniquely identify a patient.

  • Patient name: field used to identify the patient with his/her full name

  • Nickname: field used to identify the patient with his preferred name or nickname.

  • DoB - patient date of birth and current age;

  • Patient ID - patient document ID (optional field)

  • MRN: field to identify the appointment ID.

Tip

TIP - Monitoring sessions with the same MRN and DoB will be tied together in a PDF report as current and past monitoring sessions for comparison reasons.Monitoring observations:

The second section gives an overview of the patient’s clinical context and overall session results.

...

image-20240815-162844.pngImage Added

Sub-section

Image

Description

Mean estimated parameters with visual aids

...

image-20240815-162118.pngImage Added
  • mean est. P2P1 = this is the mean value of the estimated P2P2 ratio for the overall monitoring session

  • mean nTTP = this is the mean value of the normalized Time-to-Peak overall monitoring session

ATTENTION - the coloring for each threshold of parameter values is defined by your healthcare institution. This is not a diagnosis information. Please refer to our https://brain4care.atlassian.net/wiki/spaces/B4CSENSORGUIDE30/pages/3220243685/3.1+purpose+of+use#3.1.1-indications-for-use

back to top


Derived parameters graphs:

...

image-20240815-162747.pngImage Added

Sub-section

Image

Description

Session counter

...

image-20240815-170310.pngImage Added

When several sessions are tied together via same MRN and DoB, this counter will show which session this one is in that history of sessions.

Date and duration

...

image-20240815-170332.pngImage Added

Date - MM/DD/YYY of monitoring session

Duration - HH MM of monitoring session

Unique identifier

...

image-20240815-170359.pngImage Added

Unique Monitoring Identifier - 12 alphanumeric number unique to a monitoring sessions where the first 6 characters are the sensor’s serial number

Minute counter

...

image-20240815-171951.pngImage Added

Sequential minute counter for monitoring session. Every minute will have three 20s derived parameter

Derived parameters with visual aids

...

image-20240815-172103.pngImage Added

Visual aids to support interpretation of surrogate ICP waveform derived parameters.

ATTENTION - the coloring for each threshold of parameter values is defined by your healthcare institution. This is not a diagnosis information. Please refer to our https://brain4care.atlassian.net/wiki/spaces/B4CSENSORGUIDE30/pages/3220243685/3.1+purpose+of+use#3.1.1-indications-for-use

Pause / clinical notes

...

image-20240815-170937.pngImage Added

image-20240815-171032.pngImage Modified - indicates that there was a pause in the monitoring session

image-20240815-171104.pngImage Modified - indicates that a clinical note was included in the respective minute of that monitoring session

Sensor / patient position

...

image-20240815-171143.pngImage Added

Top - indicates sensor position on patient head (L = left | R = right)

Bottom - indicates patient’s body position (Standing, Sitting, or XX degrees supine)

Start / end time

image-20240815-171627.pngImage Modified

Start - left site (HH MM)

End - right side (HH MM)

Duration

image-20240815-171652.pngImage Modified

Total duration of a monitoring session.

Info

ATTENTION - These estimated P2/P1 peaks identified in this report are the most likely P1 and P2 positions derived from a statistical model based on the cardiac cycle of patients. The estimated P2/ P1 ratio shows the value of the division between the amplitude of peak P2 and the amplitude of peak P1. Values above 1.0 indicate that P2 is greater than P1 when monitored with the brain4care sensor.

Info

ATTENTION - The B4C System is an operator dependent device that contains inherent variances and as such all confidence intervals should be considered by the health professional during interpretation of estimated waveform associated parameters. 

Info

ATTENTION - The surrogate waveform and estimated associated parameters are intended to be utilized to monitor relative variation in an individual patient’s intracranial pressure over time.

Note

PRECAUTION - The derived P2/P1 ratio is an estimated value and therefore carries some uncertainty. The clinician is reminded to use their professional judgment when interpreting this and all waveform information in the report.

Warning

WARNING - The B4C System does not replace a comprehensive clinical evaluation. The waveform output should always be evaluated by the clinician in conjunction with other clinical parameters or analyses

back to top


Footer (legend and disclaimers):

The legend section repeats itself in the Summary section and also Notes and comparison section.

...

image-20240815-163110.pngImage Added

Sub-section

Image

Description

Sensor and patient position

...

image-20240815-192825.pngImage Added

Sensor position - right or left side of patient head

Patient position - help inform the patient position at the beginning of the monitoring session and if there were change during monitoring session

Visual aid legend

image-20240815-193217.pngImage Modified

Visual aid legend to support interpretation of surrogate ICP waveform derived parameters.

ATTENTION - the coloring for each threshold of parameter values is defined by your healthcare institution. This is not a diagnosis information. Please refer to our https://brain4care.atlassian.net/wiki/spaces/B4CSENSORGUIDE30/pages/3220243685/3.1+purpose+of+use#3.1.1-indications-for-use

Clinical notes

...

image-20240816-205803.pngImage Added

The clinical notes is a placeholder for any hand written clinical notes.

In future releases the clinician will be able to add them in the app itself.

The footer is repeated in all pages

...

image-20240815-163134.pngImage Added

It is included in every report page with relevant disclaimer plus

  • customer support email contact

  • customer support telephone

  • support web address

  • hospital name

  • UMI: report unique identifier

  • report page number

back to top


SECTION 2 - Notes and comparison

The second report section will give a summary of the current monitoring session with the additional information of clinical notes made during the monitoring session and also any historical monitoring sessions for longitudinal comparisons.

Tip

TIP - Monitoring sessions with the same MRN and DoB of this monitoring report will be tied together in this section as current and past monitoring sessions comparison reasons. It is most commonly utilized for longitudinal patient analysis.

Current session with notes:

...

image-20240815-193828.pngImage Added

Sub-section

Image

Description

Session counter

...

image-20240816-211115.pngImage Added

When several sessions are tied together via same MRN and DoB, this counter will show which session this one is in that history of sessions.

Date and duration

...

image-20240815-170332.pngImage Added

Date - MM/DD/YYY of monitoring session

Duration - HH MM of monitoring session

Unique identifier

...

image-20240815-170359.pngImage Added

Unique Monitoring Identifier - 12 alphanumeric number unique to a monitoring sessions where the first 6 characters are the sensor’s serial number

Mean P2P1 with visual aid

...

image-20240816-212202.pngImage Added

mean est. P2P1 = this is the mean value of the estimated P2P2 ratio for the overall monitoring session

ATTENTION - the coloring for each threshold of parameter values is defined by your healthcare institution. This is not a diagnosis information. Please refer to our https://brain4care.atlassian.net/wiki/spaces/B4CSENSORGUIDE30/pages/3220243685/3.1+purpose+of+use#3.1.1-indications-for-use

Average waveform shape

...

image-20240815-194329.pngImage Added

A picture of the average shape of the waveform during the monitoring session.

  • solid black line - overall average shape

  • blue lines - overlapping shapes for every 20s segment

Mean nTTP with visual aid

...

image-20240816-212240.pngImage Added

mean nTTP = this is the mean value of the normalized Time-to-Peak overall monitoring session

ATTENTION - the coloring for each threshold of parameter values is defined by your healthcare institution. This is not a diagnosis information. Please refer to our https://brain4care.atlassian.net/wiki/spaces/B4CSENSORGUIDE30/pages/3220243685/3.1+purpose+of+use#3.1.1-indications-for-use

Visual aid for P2P1

...

image-20240816-210738.pngImage Added

Visual aid legend to support interpretation of surrogate ICP waveform derived parameters.

ATTENTION - the coloring for each threshold of parameter values is defined by your healthcare institution. This is not a diagnosis information. Please refer to our https://brain4care.atlassian.net/wiki/spaces/B4CSENSORGUIDE30/pages/3220243685/3.1+purpose+of+use#3.1.1-indications-for-use

Visual aid for nTTP

...

image-20240816-210839.pngImage Added

Clinical notes

...

image-20240816-210929.pngImage Added

The clinical notes sub-section will list all clinical notes made during the current monitoring session

back to top


Past sessions:

...

image-20240816-211958.pngImage Added

Sub-section

Image

Description

Past session counter

...

image-20240816-211115.pngImage Added

When several sessions are tied together via same MRN and DoB, this counter will show the session’s number relative to the first session (most in the past) tied together

Date and duration

...

image-20240815-170332.pngImage Added

Date - MM/DD/YYY of monitoring session

Duration - HH MM of monitoring session

Unique identifier

...

image-20240815-170359.pngImage Added

Unique Monitoring Identifier - 12 alphanumeric number unique to a monitoring sessions where the first 6 characters are the sensor’s serial number

Average waveform shape

...

image-20240815-194329.pngImage Added

A picture of the average shape of the waveform during the monitoring session.

  • solid black line - overall average shape

  • blue lines - overlapping shapes for every 20s segment

Visual aid for P2P1

...

image-20240816-210738.pngImage Added

Visual aid legend to support interpretation of surrogate ICP waveform derived parameters.

ATTENTION - the coloring for each threshold of parameter values is defined by your healthcare institution. This is not a diagnosis information. Please refer to our https://brain4care.atlassian.net/wiki/spaces/B4CSENSORGUIDE30/pages/3220243685/3.1+purpose+of+use#3.1.1-indications-for-use

Visual aid for nTTP

image-20240816-210839.pngImage Added

...

back to top


SECTION 3 - Minute by minute results

The third report section will show a minute by minute average surrogate waveform.

...

image-20240816-213449.pngImage Added
Info

ATTENTION - you may see following messages will appear in place of the average pulse waveform and associated parameters

  • No Data - when no data for that minute was injected into the database

  • Insufficient signal quality - when less than half of the six 10 second blocks of that minute did not pass the SNR (signal-to-noise) filter

  • Insufficient sample size - at least half of expected pulses, based on derived BPM, were considered artifacts

  • Difference between P2/P1 statistically non-significant - WHEN the smallest of P2 or P1 confidence intervals is bigger than 20% of pulse’s max amplitude OR the smallest of P2 or P1 max amplitudes MINUS max between P2 or P1 min amplitudes is greater than 30% of smallest of P2 or P1 confidence intervals

  • Time to peak out of physiological limits - none of surrogate ICP pulses in that minute had a nTTP between 0 and 0.5

Sub-section

Image

Description

Past session counter

...

image-20240816-211115.pngImage Added

When several sessions are tied together via same MRN and DoB, this counter will show the session’s number relative to the first session (most in the past) tied together

Date and duration

...

image-20240815-170332.pngImage Added

Date - MM/DD/YYY of monitoring session

Duration - HH MM of monitoring session

Unique identifier

...

image-20240815-170359.pngImage Added

Unique Monitoring Identifier - 12 alphanumeric number unique to a monitoring sessions where the first 6 characters are the sensor’s serial number

Average waveform shape

...

image-20240816-213700.pngImage Added

A picture of the average shape of the waveform for a specific minute of the monitoring session

  • blue line - average shape of surrogate ICP waveform in that minute

  • first dotted vertical line - time for beginning of nTTP’s timer

  • solid black vertical line - time of normalized time to peak

  • grey block with dotted vertical line - area for estimated P1 location (left) and P2 location (right)

Derived parameters for that minute

...

Warning
image-20240816-214237.pngImage Added
  • P2/P1 - estimated P2P1 ration and confidence interval

  • nTTP - normalized time-to-peak and confidence interval

  • HR (bpm) - derived heart rate with confidence interval

  • Useful pulses - derived useful pulses as % of total identified pulses

Disclaimers

Info

ATTENTION - These estimated P2/P1 peaks identified in this report are the most likely P1 and P2 positions derived from a statistical model based on the cardiac cycle of patients. The estimated P2/ P1 ratio shows the value of the division between the amplitude of peak P2 and the amplitude of peak P1. Values above 1.0 indicate that P2 is greater than P1 when monitored with the brain4care sensor.

Info

ATTENTION - The B4C System is an operator dependent device that contains inherent variances and as such all confidence intervals should be considered by the health professional during interpretation of estimated waveform associated parameters. 

Info

ATTENTION - The surrogate waveform and estimated associated parameters are intended to be utilized to monitor relative variation in an individual patient’s intracranial pressure over time.

Note

PRECAUTION - The derived P2/P1 ratio is an estimated value and therefore carries some uncertainty. The clinician is reminded to use their professional judgment when interpreting this and all waveform information in the report.

WARNING - The B4C System does not replace a comprehensive clinical evaluation. The waveform output should always be evaluated by the clinician in conjunction with other clinical parameters or analyses\back to top