7.1 BcSs-PICNIW-1000 and BcSs-PICNIW-2000 service life
Although the BcSs-PICNIW-1000 or BcSs-PICNIW-2000 does not have a specified service life, it is recommended to change the battery every one and a half years, considering a daily and intense sensor usage.
Due to the sensitivity of the device, the equipment should be handled with care to avoid any impact with rigid objects.
Avoid contact of the sensor with knives, box cutters, or sharp materials to prevent tearing or damaging the silicone parts.
7.2 shelf life of accessories and parts
All devices in the B4C System do not have a specified service life. Due to the sensitivity of the devices, they should be handled with care to avoid any impact with rigid objects.
7.3 preventive or periodic maintenance
Periodic annual checks are recommended, which may include BcSs-PICNIW-1000 or BcSs-PICNIW-2000 internal verification and recalibration executed by brain4care support team. Such verification must be performed exclusively by brain4care authorized service.
7.4 equipment maintenance alert
If the hospital or responsible institution does not use the equipment properly through a satisfactory maintenance program, the result may be defective equipment and possible health risks.
7.5 storage instructions
Storage conditions
Info |
---|
ATTENTION - Always turn off the sensor when storing it |
All components should be stored in the product casing (Figure 7.1-1) and in a location with restricted personnel access, as well as a dust-free environment. The BcSs-PICNIW-2000 Sensor and accessories may be stored in temperatures ranging between -10 ~ 45 Celsius (14 ~ 113 Fahrenheit) and humidity ranges between 15 to 85 %.
...
Storage period
If components are stored for an extended period of time, execute a visual inspection, testing and cleaning at least once every six months for proper maintenance.
...
All relevant maintenance information are listed bellow
Child pages (Children Display) | ||
---|---|---|
|