For proper monitoring, you’ll need to properly position the sensor on the patient’s head and verify the acquired signal.

Info

ATTENTION - THIS STEP IS THESE STEPS ARE VERY IMPORTANT. PLEASE READ THEM CAREFULLY.

Anchor
top
top

Table of Contents
minLevel1
maxLevel6
include
outlinefalse
indent
styledefault
exclude
typelist
class
printablefalse

...

When starting sensor

Sensor placement

a screen to help with sensor positioning. This will inform you about the headband tightness. Before placing the sensor on the patient's head the option will show that the sensor is too loose.

SENSOR PLACEMENT

is a key part of a monitoring session.

Excerpt
name6.7 - position sensor on patient

Step 1 - position sensor

Image RemovedImage Removed
Image AddedImage Modified

POSITION image-20240815-152420.png = The sensor must be positioned in the region of the temples (mid eyebrow line), i.e., at 15% of the reference point Fpz towards the Oz reference points at the back of the head following the circumference of the skull.

POSITION image-20240815-152445.png = The sensor should contact the scalp in the temporoparietal region about 2 inches (5-6 cm) above the entrance of the external acoustic meatus in the coronal plane and just below the head protrusion where the circumference is greatest.

back to top


Step 2 - tighten turnbuckle

Once the sensor is placed, tighten the turnbuckle into a snug, comfortable fit, turning it clockwise until a final click can be heard without excessive force. A proper adjustment will result in a green OK! on the app screen.

Image RemovedImage Added
  • If the patient is awake, ask if he is comfortable with the pressure. If not, readjust the headband for a comfortable fit.

  • In conditions where patients are unresponsive, check for signs of discomfort or skin damage for at least every five minutes.

...

Sensor too Lose

Sensor too tight

Excessive tightening will be indicated with a “sensor too tight” message. Loosen the turnbuckle until the green message of OK! appears.

Lack of tightening will be indicated with a “sensor too lose” message. Tighten the turnbuckle until the green message of OK! appears.

Image RemovedImage AddedImage RemovedImage Added
Note

PRECAUTION - The patient position or movement and the BcSs-PICNIW-1000/BcSs-PICNIW-2000 sensors are known to affect the signal being recorded. As a result, if multiple recording sessions are obtained on the same patient, consistent positioning of the patient and device will lead to more comparable recordings.

...

After sensor placement and tightening an assisted waveform quality window will appear via sample processing according to the table below.

Image RemovedImage Added

Sample gathering - when correctly positioned, the B4C System will start collecting 20s or 30s sample for quality verification.

Image RemovedImage Added

Adequate quality - An average pulse will appear with respective confidence intervals. It will refresh every 5s.

Image RemovedImage Added

Inadequate quality - In some situations, the captured signal does not have sufficient technical quality.

Verify if the patient is quiet and/or adjust the sensor.

...