CONTRAINDICATION - Do NOT use this device on patients without cranial bone flap or have extensive scalp injury, bone defects (sections of skull missing), or any other conditions that deems to be a contraindication to performing monitoring using the brain4care device.
The
...
optimal position for ICP waveform monitoring is
...
with the patient in a
...
supine position, angled at 30 degrees, with the chin and sternum aligned (
...
see figures below).
...
Ensure the head
...
is not
...
tilted
...
and the patient remains still to avoid interference with the ICP waveform reading.
...
Note |
---|
PRECAUTION - The patient position or movement and the BcSs-PICNIW-1000 |
...
and BcSs-PICNIW-2000 sensors is known to affect the signal being recorded. As a result, if multiple recording sessions are obtained on the same patient, consistent positioning of the patient and device will lead to more comparable recordings. |