Step 1 (option A) - turn on
...
sensor BcSs-PICNIW-2000
To turn on the sensor to initiate pairing procedure. Verify if sensor is on via LED indication. If not, long press the on/off button util the LED is on.
...
Turn on sensor using power switch, setting it to “I” position.
...
Info |
---|
ATTENTION - The maximum operating temperature of the BcSs-PINICW-2000 sensor model is 35°C. Therefore, if it is stored in a location of up to 40°C, it is recommended to use the sensor in a place with a temperature within the operating range (0 to 35°C) and wait until the sensor cools down. If the operating environment is at 20°C, the sensor takes approximately 6 minutes to cool down and reach the operating temperature (35°C). |
After 5 minutes of inactivity (not being used or not on the charging cradle) the sensor will turn off.
To turn on the sensor to initiate the pairing procedure by using the power switch, setting it to “I” position.
Make sure you are not trying the press the MODE button.
This is used to pair the sensor with a dongle (please see section https://brain4care.atlassian.net/wiki/spaces/
...
...
3220246611/6.8+multiparameter+monitor+optional#6.8.1-manual-pairing if you are using a dongle).
Info |
---|
ATTENTION - If the switch is not in its final position, the internal pin protector may interfere with adequate pin movement, decreasing sensor sensitivity, or even causing misreadings. |
Verify if sensor is on via LED indication. If not, long press the on/off button util the LED is on.
...
...
automatic sensor-app pairing is a default system feature unless an organization has more than one active sensor. In this case automatic sensor pairing will happen only for the most recently used sensor. |
Click on the connect button (1). The app will scan for any close by sensors that are ON.
Step 3 - confirm sensor pairing
A list of sensors will appear on
...
the sensor pairing pop-up screen ( 1 ). The closest sensor will be selected automatically. Verify
...
that the
...
UDI (Unique Device ID) on the sensor, located opposite the power buttons,
...
matches the sensor selected in the mobile app.
Tip |
---|
TIP - Some sensors may not be available for selection (dimmed) because they are blocked or not active for your organization. |
A visual confirmation for pairing will flash. The sensor and
...
its image in the mobile app will blink with the same color and frequency ( 2 ). If
...
the blinking pattern matches between the mobile app
...
and the sensor, the
...
sensor is
...
correctly selected. Click
...
"Confirm" ( 3 ). If the
...
patterns do not match or if you
...
wish to select a different sensor, click "Cancel" ( 4 ). Once
...
confirmed, the sensor’s status LED
...
will flash green three times.