Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

7.1 bcss-picniw-1000 service life

Although the BcSs-PICNIW-1000 does not have a specified service life, it is recommended to change the battery every one and a half years, considering a daily and intense sensor usage.
Due to the sensitivity of the device, the equipment should be handled with care to avoid any impact with rigid objects.

7.2 shelf life of accessories and parts

All devices in the B4C System do not have a specified service life. Due to the sensitivity of the devices, they should be handled with care to avoid any impact with rigid objects.

7.3 preventive or periodic maintenance

Periodic annual checks are recommended, which may include BcSs-PICNIW-1000 internal verification and recalibration executed by brain4care support team. Such verification must be performed exclusively by brain4care authorized service.

7.4 equipment maintenance alert

If the hospital or responsible institution does not use the equipment properly through a satisfactory maintenance program, the result may be defective equipment and possible health risks.

7.5 storage instructions

Storage conditions

Info

ATTENTION - Always turn off the sensor when storing it

All components should be stored in the product casing (Figure 7.1-1) and in a location with restricted personnel access, as well as a dust-free environment. The BcSs-PICNIW-2000 Sensor and accessories may be stored in temperatures ranging between -10 ~ 45 Celsius (14 ~ 113 Fahrenheit) and humidity ranges between 15 to 85 %.

...

Storage period

If components are stored for an extended period of time, execute a visual inspection, testing and cleaning at least once every six months for proper maintenance.

...

All relevant maintenance information are listed bellow

Child pages (Children Display)
allChildrentrue