SECTION 1 - Summary

The report generated by the B4C System has five sections that organize all information acquired and processed for a monitoring session. Refer to 12 technical documentation for further details regarding the analytical software.

...

back to top

...

Report Header

The first report section is a summary of the monitoring session information including patient identification, monitoring session identification, sensor and patient information, observations, and patient clinical context.

...

Patient Identification:

this sub-section contains all information to uniquely identify a patient.

...

  • support email contact

  • support telephone

  • support web address

  • hospital name

  • UMI: report unique identifier

  • report page number

...

back to top

...

SECTION 2 - Surrogate ICP waveform report

The second report section is a summary of the monitoring session pulse morphology and relevant disclaimers.

...

Contains important information and disclaimers that should be read by the clinician.

...

back to top

...

SECTION 3 - Monitoring observations

The third report section appears only for monitoring sessions that have had an observation included, will show the observation details for each minute that has an observation. It can be one to several pages depending on how many observations were made.

...

Minute’s observations: this box contains the actual observations included for that minute. There can be a maximum of 6 observations. Each observation is numbered and includes the sensor and patient position informed by the operator at the moment of observation.

...

back to top

...

SECTION 4 - Minute by minute session results

The fourth report section will show a minute by minute average surrogate waveform calculation for the whole monitoring session. Each subsection will have two swim-lanes for each 4 minutes of monitorization. It could be one to many pages depending on how many minutes the monitoring session lasted.

...

Info

ATTENTION - you may see following messages will appear in place of the average pulse waveform and associated parameters

  • No Data - when no data for that minute was injected into the database

  • Insufficient signal quality - when less than half of the six 10 second blocks of that minute did not pass the SNR (signal-to-noise) filter

  • Insufficient sample size - at least half of expected pulses, based on derived BPM, were considered artifacts

  • Difference between P2/P1 statistically non-significant - WHEN the smallest of P2 or P1 confidence intervals is bigger than 20% of pulse’s max amplitude OR the smallest of P2 or P1 max amplitudes MINUS max between P2 or P1 min amplitudes is greater than 30% of smallest of P2 or P1 confidence intervals

  • Time to peak out of physiological limits - none of surrogate ICP pulses in that minute had a nTTP between 0 and 0.5

...

back to top

...

SECTION 5 - Additional information

The fifth report section which appears in the last page, contains more technical information for the entire monitoring session. It contains five graphs as explained below. This is typically used in researches or customer support context.

...