The B4C System is designed to provide for non-invasive monitoring of variation in intracranial pressure (ICP) variations. The system ONLY It monitors only the intracranial pressure ( ICP ) pulse morphology, which is shown displaying it as a surrogate ICP waveform on the brain4care application or, with the use of the optional dongle, show the sensor waveform in on a compatible multiparameter monitor that . The monitor must have an Invasive Blood Pressure (IBP) or Invasive Pressure (IP) Module with sensitivities of 5 μV/Vex/mmHg or higher and have automatic amplitude window adjustment.
Important:
This system does NOT
...
measure actual pressure in mmHg.
The signal
...
displayed on the mobile application or patient monitor should be
...
interpreted based solely on its waveform morphology, not its absolute ICP value.
The BcSsPICNIW-1000
...
Sensor and B4C System
...
are not intended
...
to be used as
...
standalone diagnostic
...
tools. The surrogate ICP waveform
...
should be considered as one element
...
within a
...
broader clinical evaluation.
Info |
---|
ATTENTION - This sensor does NOT monitor measure ICP in mmHg and therefore . Therefore, the waveform shown in a displayed on the monitor should be assessed only solely by its morphology and NOT , not by its absolute value. |
Anchor | ||||
---|---|---|---|---|
|
...