The B4C System is designed to provide for non-invasive monitoring of variation in intracranial pressure (ICP) variations. The system ONLY It monitors only the intracranial pressure ( ICP ) pulse morphology, which is shown displaying it as a surrogate ICP waveform on the brain4care application or, with the use of the optional dongle, show the sensor waveform in on a compatible multiparameter monitor that . The monitor must have an Invasive Blood Pressure (IBP) or Invasive Pressure (IP) Module with sensitivities of 5 μV/Vex/mmHg or higher and have automatic amplitude window adjustment.

Important:

  • This system does NOT

...

  • measure actual pressure in mmHg.

  • The signal

...

  • displayed on the mobile application or patient monitor should be

...

  • interpreted based solely on its waveform morphology, not its absolute ICP value.

  • The BcSsPICNIW-1000

...

  • Sensor and B4C System

...

  • are not intended

...

  • to be used as

...

  • standalone diagnostic

...

  • tools. The surrogate ICP waveform

...

  • should be considered as one element

...

  • within a

...

  • broader clinical evaluation.

Info

ATTENTION - This sensor does NOT monitor measure ICP in mmHg and therefore . Therefore, the waveform shown in a displayed on the monitor should be assessed only solely by its morphology and NOT , not by its absolute value.

Anchor
top
top

...