Figure 6.1-2a Detaching components BcSs-PICNIW-1000 sensor

image-20240206-220737.png

Figure 6.1-2b Detaching components BcSs-PICNIW-2000 sensor

back to top

...

Step 3 - apply alcohol

Apply ethyl or isopropyl alcohol (70-90%) to sterile gauze towels and gently wipe each component (sensor, battery, headband, and turnbuckle).

back to top

...

Step 4 - let it dry

Wait for the components to dry for a few minutes and then store them in the storage case.

...

Info

ATTENTION - Disinfect the sensor and headband with 70% ethanol alcohol, or other agents according to your healthcare facility’s disinfection protocols. This must be done before and after any use on patients.

back to top

...

Step 5 - visual inspection

If the device is determined not to be visually clean and free of unacceptable deterioration at the end of the cleaning step, either repeat the relevant previous cleaning steps or safely dispose of the device.