To disinfect the sensor generously apply 70% ethanol alcohol to a sterilized gauze and wipe each component according to the following instructions:

Table of Contents
minLevel1
maxLevel6
include
outlinefalse
indent
stylenone
exclude
typelist
class
printablefalse

...

Info

ATTENTION - Do NOT sterilize in an autoclave. This will damage the equipment.

Info

ATTENTION - Do NOT press the pin on the green base of the sensor during sterilization to avoid damaging the sensor.

Step 1 - turn off sensor

Before you begin, make sure the sensor is turned off verifying that the power switch set to “O”.

Figure 6.1-1a Turning off the sensor BcSs-PICNIW-1000

Before you begin, make sure the sensor is turned off by checking if the LED is turn-on. If yes, keep pressed util the LED is turn-off

Figure 6.1-1b Turning off the sensor BcSs-PICNIW-2000

...