The B4C System is designed to provide non-invasive monitoring of variation in intracranial pressure (ICP). The system ONLY monitors the intracranial pressure (ICP) pulse morphology, which is shown as a surrogate ICP waveform on the brain4care application or, with the use of the optional dongle, show the sensor waveform in a compatible multiparameter monitor that have Invasive Blood Pressure (IBP) or Invasive Pressure (IP) Module with sensitivities of 5 μV / Vex / mmHg or higher and have automatic amplitude window adjustment. This system does NOT monitor actual pressure in mmHg. The signal shown on the mobile application or patient monitor should be assessed only by its morphology and NOT by its ICP absolute value. The BcSsPICNIW-1000 / Sensor and B4C System is not intended for use as a standalone diagnostic tool. The surrogate ICP waveform is one element of a comprehensive clinical evaluation.

...