...
Sensor too Lose | Sensor too tight |
---|---|
Excessive tightening will be indicated with a “sensor too tight” message. In such a case loosen Loosen the turnbuckle until the green message of OK! appears. | Lack of tightening will be indicated with a “sensor too lose” message. In such a case tighten Tighten the turnbuckle until the green message of OK! appears. |
Note |
---|
PRECAUTION - The patient position or movement and the BcSs-PICNIW-1000/BcSs-PICNIW-2000 sensors are known to affect the signal being recorded. As a result, if multiple recording sessions are obtained on the same patient, consistent positioning of the patient and device will lead to more comparable recordings. |
...
After sensor placement and tightening an assisted waveform quality window will appear via sample processing according to the table below.
Sample gathering - when correctly positioned, the B4C System will start collecting |
20s or 30s sample for quality verification. | |
Adequate quality - An average |
pulse will |
appear with respective confidence intervals |
. It will |
refresh every |
5s. | |
Inadequate quality - In some situations, the captured signal does not have sufficient technical quality. |
Verify if the patient is quiet and/or adjust the sensor. |
Note |
---|
PRECAUTION - The signal quality check is NOT a clinical judgement or diagnosis. The is a technical verification and the insufficient signal quality message means one or more of the four conditions below are true:
|
If the sensor is positioned correctly, the waveform shown on the positioning screen should have a typical ICP waveform characteristic.
Info |
---|
ATTENTION - The familiarity with the waveform is a healthcare professional interpretation based on their experience and exposure to ICP waveforms. |
...