Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Sensor too Lose

Sensor too tight

Excessive tightening will be indicated with a “sensor too tight” message. In such a case loosen Loosen the turnbuckle until the green message of OK! appears.

Lack of tightening will be indicated with a “sensor too lose” message. In such a case tighten Tighten the turnbuckle until the green message of OK! appears.

Note

PRECAUTION - The patient position or movement and the BcSs-PICNIW-1000/BcSs-PICNIW-2000 sensors are known to affect the signal being recorded. As a result, if multiple recording sessions are obtained on the same patient, consistent positioning of the patient and device will lead to more comparable recordings.

...

After sensor placement and tightening an assisted waveform quality window will appear via sample processing according to the table below.

Image Modified

Sample gathering - when correctly positioned, the B4C System will start collecting

a 30 second sample of the captured waveform

20s or 30s sample for quality verification.

Image Modified

Adequate quality - An average

30 second

pulse will

be presented

appear with respective confidence intervals

and

. It will

be updated

refresh every

5 seconds

5s.

Image Modified

Inadequate quality - In some situations, the captured signal does not have sufficient technical quality.

The system will indicate this, meaning there is a need to stabilize

Verify if the patient is quiet and/or adjust the sensor.

Note

PRECAUTION - The signal quality check is NOT a clinical judgement or diagnosis. The is a technical verification and the insufficient signal quality message means one or more of the four conditions below are true:

  • The signal-to-noise ratio is not satisfactory (signal with lack of physiological characteristics)

  • The amount of artifacts is excessive (too much movement being captured in the signal)

  • The processed signal will result in a statistically non-significant P2/P1 ratio

  • Processed signal will result in a non-physiological Time-to-Peak range

If the sensor is positioned correctly, the waveform shown on the positioning screen should have a typical ICP waveform characteristic.

Info

ATTENTION - The familiarity with the waveform is a healthcare professional interpretation based on their experience and exposure to ICP waveforms.

...