Turn on sensor using power switch, setting it to “I” position.

Turning ON sensor BcSs-PICNIW-1000

Make sure you are not trying the press the MODE button. This is used to pair the sensor with a dongle (please see section https://brain4care.atlassian.net/wiki/spaces/B4CSENSORGUIDE30/pages/3220246611/6.8+multiparameter+monitor+optional#6.8.1-manual-pairing if you are using a dongle)

Info

ATTENTION - If the switch is not in its final position, the internal pin protector may interfere with adequate pin movement, decreasing sensor sensitivity, or even causing misreadings.

Verify if sensor is on via LED indication. If not, long press the on/off button util the LED is on.

Turning ON the sensor BcSs-PICNIW-1000

Info

ATTENTION - The maximum operating temperature of the BcSs-PICW-2000 sensor model is 35°C. Therefore, if it is stored in a location of up to 40°C, it is recommended to use the sensor in a place with a temperature within the operating range (0 to 35°C) and wait until the sensor cools down. If the operating environment is at 20°C, the sensor takes approximately 6 minutes to cool down and reach the operating temperature (35°C).

...