Turn on sensor using power switch, setting it to “I” position. Turning ON sensor BcSs-PICNIW-1000 Make sure you are not trying the press the MODE button. This is used to pair the sensor with a dongle (please see section https://brain4care.atlassian.net/wiki/spaces/B4CSENSORGUIDE30/pages/3220246611/6.8+multiparameter+monitor+optional#6.8.1-manual-pairing if you are using a dongle) Info |
---|
ATTENTION - If the switch is not in its final position, the internal pin protector may interfere with adequate pin movement, decreasing sensor sensitivity, or even causing misreadings. |
| Verify if sensor is on via LED indication. If not, long press the on/off button util the LED is on. Turning ON the sensor BcSs-PICNIW-1000 Info |
---|
ATTENTION - The maximum operating temperature of the BcSs-PICW-2000 sensor model is 35°C. Therefore, if it is stored in a location of up to 40°C, it is recommended to use the sensor in a place with a temperature within the operating range (0 to 35°C) and wait until the sensor cools down. If the operating environment is at 20°C, the sensor takes approximately 6 minutes to cool down and reach the operating temperature (35°C). |
|
...