Equipment type BF. Continuous operation. Internal and external power through batteries only. Risk Class II, in accordance with CRF Title 21. It accompanies a protection class II power adapter and charger.
Table 3.4-1 - Equipment Classification
Equipment classification in accordance with applicable standards | |
---|---|
Protection against electric shock | Internally powered equipment |
Degree of protection against electric shock | Applied Part – Type BF |
Degree of protection against harmful water penetration | IPX0 (no protection against harmful water penetration) |
Sterilization or disinfection methods | Sterilization not applicable. Disinfection with 70% alcohol or 2% chlorhexidine |
Degree of application safety in the presence of FLAMMABLE ANESTHETIC MIX | Not suitable for use in the presence of FLAMMABLE ANESTHETIC MIXTURE with air, oxygen or nitrous oxide |
Operation mode | Continuous Operation Equipment |
...