Step 1 - turn on the sensor

To turn on the sensor to initiate pairing procedure.

Turn on sensor using power switch, setting it to “I” position.

Turning ON sensor BcSs-PICNIW-1000

Make sure you are not trying the press the MODE button. This is used to pair the sensor with a dongle (please see section https://brain4care.atlassian.net/wiki/spaces/B4CSENSORGUIDE29/pages/2853211499#6.8.1-manual-pairing if you are using a dongle)

Info

If the switch is not in its final position, the internal pin protector may interfere with adequate pin movement, decreasing sensor sensitivity, or even causing misreadings.

Verify if sensor is on via LED indication. If not, long press the on/off button util the LED is on.

Turning ON the sensor BcSs-PICNIW-1000

Info

ATTENTION - The maximum operating temperature of the BcSs-PICW-2000 sensor model is 35°C. Therefore, if it is stored in a location of up to 40°C, it is recommended to use the sensor in a place with a temperature within the operating range (0 to 35°C) and wait until the sensor cools down. If the operating environment is at 20°C, the sensor takes approximately 6 minutes to cool down and reach the operating temperature (35°C).

...

Step 2 - start sensor pairing

Click on the (plus) connect button (1). The app will scan for any close by sensors that are ON.

...

Step 3 - confirm sensor

A list of sensors will appear on your sensor pairing pop-up screen (1). The closest sensor will be selected automatically. Verify if the sensor UDI (Unique Device ID) on the sensor, opposite the power buttons, is the same that has been selected from the sensor list in the mobile app.

...