Although the BcSs-PICNIW-1000 or nor BcSs-PICNIW-2000 does do not have a specified service life, it is recommended to change the battery every one and a half years, considering a daily and intense sensor usage.
Due to the sensitivity of the device, the equipment should be handled with care to avoid any impact with rigid objects.

...

Periodic annual checks are recommended, which may include BcSs-PICNIW-1000 or BcSs-PICNIW-2000 internal verification and recalibration executed by brain4care support team. Such verification must be performed exclusively by brain4care authorized service. Refer to 9 technical support section for contact information.

7.4 equipment maintenance alert

...

All components should be stored in the product casing (Figure 7.1-1) and in a location with restricted personnel access, as well as a dust-free environment. The

BcSs-PICNIW-2000

...

BcSs-PICNIW-1000

image-20240221-201954.pngImage AddedImage Added

Storage Temperature 32 °F to 104 °F

Storage Temperature 14 °F to 113 °F

Storage period

If components are stored for an extended period of time, execute a visual inspection, testing and cleaning at least once every six months for proper maintenance.

...