A satisfactory waveform reading may present any of the typical ICP shape characteristics below according to the patient’s clinical condition (from normal to pathological)

As a suggestive reference, the waveforms below are non invasive ICP waveforms that can be shown with the noninvasive ICP sensor.

Warning

WARNING - The appearance of a normal waveform morphology does not eliminate the possibility of elevated intracranial pressure or irregularity in brain compliance.

 

Info

ATTENTION - The waveforms above are examples of typical noninvasive ICP waveforms in different clinical settings. You may see variations in the ICP waveform, but they should always follow a similar pattern.

The monitoring session should be as long as required to assess the ICP waveform. Typical sessions are of 15+ minutes but the final duration is ultimately defined by a trained professional. The session should be finalized upon patient request due to discomfort or decision by the primary caregiver.

The ICP pulse waveform (Figure 6.6-9) is an alternative to assess certain ICP characteristics, irrespective of the absolute values that are typically captured by an invasive method .

Image Removedimage-20240221-184829.pngImage Added

The ICP waveform contains characteristics that can be mathematically analyzed such as:

  • Wave peaks: P1 (percussion wave, due to arterial pressure being transmitted from the choroid plexus to the brain ventricles), P2 (tidal wave, related to brain compliance) and P3 (dicrotic wave);

  • Absolute amplitude of wave peaks: dP1 (relative amplitude of percussion wave), dP2 (relative amplitude of tidal wave), dP3 (relative amplitude of dicrotic wave);

  • Absolute lags in milliseconds of wave peaks: LT (considered the initial timestamp of a pulse), L1 (lag for percussion wave), L2 (lag for tidal wave), L3 (lag for dicrotic wave), Lx (variable lag);

  • Absolute curvature of wave peaks: Curv1 (percussion wave curvature), Curv2 (tidal wave curvature), Curv3 (dicrotic wave curvature).

Under normal ICP conditions, the relative amplitudes of the three peaks are related as follows: P1 > P2 > P3 (Box A in Figure above). In conditions with decreased brain compliance and rising ICP, the pulse waveform morphology gradually changes and certain indicators, such as the estimated P2/P1 ratio eventually increases.

Warning

WARNING - The BcSs-PICNIW-1000 Sensor does not replace a comprehensive clinical evaluation. The waveform output should always be evaluated by the clinician in conjunction with other clinical parameters or analyses.

Warning

WARNING - The B4C System has not been validated to be a substitute for invasive ICP monitoring methods in any cases.

Warning

WARNING - The medical professional is responsible for determining the additional clinical parameters or analyses required in addition to the BcSs-PICNIW-1000 Sensor waveform to make a clinically informed decision.

Warning

WARNING - The appearance of a normal waveform morphology does the eliminate the possibility of elevated intracranial pressure or irregularity in brain compliance.

Warning

WARNING - The clinician is reminded to assess the waveform morphology in real-time during a monitoring session. The B4C report is available for the clinician’s convenience to review associated waveform parameter estimates and other clinical observations entered in a past monitoring session