Designed and manufactured in accordance with ANSI/AAMI ES60601IEC 60601-1 :2005/(R)2012 And A1:2012 + Emend 1 (2016Ed. 3.0 (2005) + Amd.1 (2012) + Amd.2 (2020) and its side standards IEC 60601-1-2 : 2014, ABNT NRB IEC 60601-1-6:2011 (identical to Ed. 4.0 (2014) + Amd.1 (2020), IEC 60601-1-6: Ed.1.1b:2013), ABNT NRB IEC 60601-1-9:2014 (identical to 2019, IEC 60601-1-9 ED. 1.1 B:2013 ) and ABNT NBR IEC 60601-1-11 :2012 (identical to IEC 60601-1-11:2010).

Equipment tested under specification FCC 47 CFR Part 15 Subpart C, §15.209 Radiated emission limits and General requirements RSS-GEN, Issue 4, Nov.2014, section 8.9 Contains FCC ID: 2AA9B04

Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.Ed. 2.0 (2015) + Amd.1 (2020).

Additional details can be requested at brain4care (www.brain4.care).

3.4.1 biocompatibility

The module modules BcSs-PICNIW-1000 and BcSs-PICNIW-1000 is are registered with Anvisa and has have passed biocompatibility tests in accordance with NBR IEC 60601-1, standards ISO 10993-10: 2010, EN ISO 10993-10: 2013 and ISO 10993-5: 2009.

3.4.2 classification

Equipment , Type type BF. Continuous operation. Internal and external power through batteries only. Risk Class I, in accordance with CRF Title 21. It accompanies a protection class II power adapter and charger.

Table 3.4-1 - Equipment Classification

Equipment classification in accordance with applicable standards

Protection against electric shock

Internally powered equipment

Degree of protection against electric shock

Applied Part – Type BF

Degree of protection against harmful water penetration

IPX0 (no protection against harmful water penetration)

Sterilization or disinfection methods

Sterilization not applicable. Disinfection with 70% alcohol or 2% chlorhexidine

Degree of application safety in the presence of FLAMMABLE ANESTHETIC MIX

Not suitable for use in the presence of FLAMMABLE ANESTHETIC MIXTURE with air, oxygen or nitrous oxide

Operation mode

Continuous Operation Equipment

3.4.3 telecommunication

Equipment tested under specification FCC 47 CFR Part 15 Subpart C, §15.209 Radiated emission limits and General requirements RSS-GEN, Issue 4, Nov.2014, section 8.9 Contains FCC ID: 2AA9B04

Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.